Improvements to the X-ray Spectrometer at the Aerosol Laboratory, Instituto de Física, UNAM

Show simple item record

dc.contributor.author . Mejia-Ponce, L. V
dc.contributor.author Hernandez-Lopez, A. E.
dc.contributor.author Reynoso-Cruces, S.
dc.contributor.author Pineda, J. C.
dc.contributor.author Mendoza-Flores, J. A.
dc.contributor.author Miranda, J.
dc.date.accessioned 2022-04-15T10:53:11Z
dc.date.available 2022-04-15T10:53:11Z
dc.date.issued 2018-08-06
dc.identifier.issn 2321-8649
dc.identifier.issn 2321-9289
dc.identifier.uri http://dspace.chitkarauniversity.edu.in/xmlui/handle/123456789/375
dc.description.abstract Due to the demands of better (accurate and precise) analytical results using X-ray Fluorescence (XRF) at the Aerosol Laboratory, Instituto de Física, UNAM, it was necessary to carry out improvements in instrumentation and analytical procedures in the x-ray spectrometer located in this facility. A new turbomolecular vacuum system was installed, which allows reaching the working pressure in a shorter time. Characteristic x-rays are registered with a Silicon Drift Detector, or SDD, (8 mm thick Be window, 140 eV at 5.9 keV resolution), working directly in a high-vacuum, permitting the detection of x-rays with energies as low as 1 keV (Na Ka) and higher counting rates than in the past. Due to the interference produced by the Rh L x-rays emitted by the tube normally used for atmospheric and food analysis with Cl K x-rays, another tube with a W anode was mounted in the spectrometer to avoid this interference, with the possibility to select operation with any of these tubes. Examples of applications in atmospheric aerosols and other samples are presented, to demonstrate the enhanced function of the spectrometer. Other future modifications are also explained. en_US
dc.language.iso en en_US
dc.relation.ispartofseries ;CHAENG/2013/51628
dc.subject Standard Reference Material 2783 en_US
dc.subject chemical composition of atmospheric aerosol en_US
dc.subject X-ray fluorescence analysis en_US
dc.subject Silicon DriftDetector SDD, en_US
dc.title Improvements to the X-ray Spectrometer at the Aerosol Laboratory, Instituto de Física, UNAM en_US
dc.type Article en_US